Assessment of the space radiation environment is a high priority for space exploration. Dr. Vincent Pisacane is working to develop a portable, low-power radiation instrument that can make real-time measurements of radiation environments to assess and reduce risk, warn a crew during the onset of enhanced radiation, and allow a crew to determine safe locations during enhanced radiation. This device could also be used by scientists to validate and improve models for the space radiation environment and to test the effectiveness of shielding materials.
Overview
Dosimetry: Microdosimeter Instrument System Suitable for Spaceflight
Principal Investigator:
Vincent L. Pisacane, Ph.D.
Organization:
United States Naval Academy
Technical Summary
Original Aims
- Demonstrate that a small, compact and portable flight-qualifiable, solid-state microdosimeter can be developed to measure quantitative information on the dose and dose distribution of energy deposited in silicon cells of tissue size and by inference in tissue.
- Analyze data from radiation beam experiments and compare with radiation transport codes to provide quantitative information on the radiation environment, potential risk and the accuracy of the codes to correctly calculate energy deposition spectra.
- With data from radiation beam experiments correlated with radiation transport codes, to determine the effectiveness of selected materials to minimize the total risk from primary and secondary radiation.
The specific objectives of the MIcroDosimeter iNstrument (MIDN) instrument are to:
- Make real-time measurements of the radiation environment to assess risk (dose equivalent).
- Actively warn crew during onset of enhanced radiation events.
- Allow crew to determine safe locations during enhanced radiation events.
- Provide observations to validate and improve space radiation environment models.
- Provide observations to validate and improve radiation transport theories for shield materials and different tissue types.
While not part of this proposal, a student design effort developed an early version of a MIDN instrument that was launched on the MidSTAR-I spacecraft in 2007 with only a short time available for its design and development by the students.
We have satisfied, most but not all, of our aims and instrument development objectives.
We successfully evolved two sets of instrumentation: a bench-top system to evaluate instrument components without regard for power or size and two prototype flight instruments. Each instrument consists essentially of a sensor, sensor electronics, amplifiers, analog-to-digital conversion and a multichannel analyzer under computer or microprocessor control. We tested these instruments at the NASA Space Radiation Laboratory (NSRL) at Brookhaven National Laboratory (BNL). Ions examined include iron, oxygen, silicon, hydrogen, carbon and titanium. In our benchtop system with a 10 um thick sensor, we were able to achieve a dE/dx < 1keV/um in silicon that is equivalent to a lineal energy of approximately 0.4 keV/um in tissue. In our flight prototype instrument with a 10 um thick sensor, we were able to achieve a dE/dx ~ 3 keV/um in silicon that is equivalent to a lineal energy of ~ 1 keV/um in tissue. The anticipated flight system will require a power of approximately one watt, could be packaged into a volume of less than 1284 cm, and should have a dE/dx < 1keV/um in silicon that is equivalent to a lineal energy of approximately 0.4 keV/um in tissue. These are significant accomplishments that satisfy the primary objectives of the research and verify our original hypotheses that silicon microdosimetery appears to be a viable alternative to assess a mixed and unknown time-varying radiation field to estimate regulatory risk.
This is the final year of this NSBRI grant.
Earth Applications
The use of prior methods is limited in part because of the complexity, sensitivity and lack of reliability of the most commonly used instruments, gas proportional counters. The compact system that we have developed for space applications would likewise be applicable for these situations and measurements described in the previous paragraph.
We have established for the first time in a solid-state microdosimeter a lowered energy cutoff of dE/dx < 1 keV/um in silicon that is equivalent to a lineal energy cutoff of < 0.4 keV/um in tissue. Thus we have an instrument that can be used in space and terrestrially to directly assess regulatory risk.